VTK-VS-BA-015e-A | User Manual MDS 3200+ Series
|
Cleaning
91
Step 6:
For fine purification you have to clean all components by hand. In general you
should take special care about those places, which get in contact with the medi-
um or where two different parts fit together. First you should clean the nozzle in-
sert and the tappet guidance, since hardening would cause most problems here.
IMPORTANT INFORMATION! (simpler cleaning)
For some uncomplicated media it is possible to clean the parts just by using a fluid like etha-
nol and compressed air, which can simplify the cleaning process. But before you try it, you
should always contact our Technical Support.
−
Start with the nozzle insert. Clean it thoroughly from above and below with a cleaning
rod. For the bore use a nozzle insert cleaning wire. These come with different sizes. They
are not part of the cleaning toolkit CTK and have to be ordered separately.
−
Use a cleaning rod for the upper part of the tappet guidance. For the rest of the outer
part of the tappet guidance use a fluidic brush and a fluid box cleaner.
−
Clean the bore of the tappet guidance with a fluidic brush. Move it back and forth several
times to clear the hole of any traces of the medium.