ERROR MESSAGES
The following pages list all the error messages you could encounter when using a MDS 3200
+
system. In case of an error, an error message is shown in the display and the red adjust LED is
switched “ON“. Press
[enter]
to clear the display. If this does not work, switch “OFF“ the control
unit. Check the system for potential problems
(e.g. loose cables)
and re-start the system.
IMPORTANT NOTE! (Storing error messages)
After the re-start, the error message can only be found in the submenu “Error“
.
If the error still occurs, please contact the Technical Support at VERMES Microdispensing or your
local supplier.
IMPORTANT NOTE! (valve status depending on error)
If the valve is open or closed after an error message, depends on the respective error. In the fol-
lowing list it is mentioned for each error.
1
Valve Error 101
This message sometimes appears during startup of the system. The control unit checks the
data of valve and control unit. If the control unit does not recognize the valve resp. the re-
quired valve type
(e.g. finds a low viscosity valve)
, the error message is displayed.
Error code display:
Valve Error 101
Error code status menu:
Valve Error 101
Error handling:
•
The sensor connection has not been plugged in correctly.
Switch OFF control unit, and inspect the connection.
If the cable turns out to be damaged, it must be exchanged
before restarting.
•
Maybe an inadequate valve type has been installed.
Valve:
Valve still open
(start-up of the MDC)
2
Valve Error 102
The piezoelectric element does not match the current application
(Valve Passport)
. This error
message appears with the start-up of the control unit.
Error code display:
Valve Error 102
Error code status menu:
Valve Error 102
Error handling:
•
The sensor connection has not been plugged in correctly.
Switch OFF control unit, and inspect the connection.
If the cable turns out to be damaged, it must be exchanged
before restarting.
•
Exchange valve
Valve:
Valve still open
(start-up of the MDC)
3
Valve Error 104
Checking the actuator temperature returns a value outside of the acceptable limits. This
error can occur anytime while the control unit is on.
Error code display:
Valve Error 104
Error code status menu:
Valve Error 104
Error handling:
•
The sensor connection has not been plugged in correctly.
Switch OFF control unit, and inspect the connection.
If the cable turns out to be damaged, it must be exchanged
before restarting.