64
VTK-VS-BA-015e-A | User Manual MDS 3200+ Series
|
Communication Interfaces
27
SYSTEM:SINGLEDOSOK:SETUP
Description:
This command sets the Single-DOSOK signal to “Setup”. The length of the
Single-DOSOK signal is that of the setup.
Example:
Input:
SYSTEM:SYNGLEDOSOK:SETUP
Result:
The single-DOSOK is set to “setup”.
Return:
OK
28
SYSTEM:SINGLEDOSOK:PULSE
Description:
This command sets the Single-DOSOK signal to “Pulse”. The length of the
Single-DOSOK signal is that of a pulse.
Example:
Input:
SYSTEM:SYSTEMDOSOK:PULSE
Result:
The single-DOSOK is set to “pulse”.
Return:
OK
29
TEMP:? (TEMP = temperature)
Description:
The command provides information about the current temperature of the
nozzle heater; the measuring unit is °C. The MDC will answer “no Heater”, if
the heater is turned “OFF” in the submenu “Heater”.
Example:
Input:
TEMP:?
Result:
70
… the current temperature value amounts to 70 °C.
Return:
70
30
TEMP:<Setpoint in °C> (TEMP = temperature)
Description:
This command is used to select a new target value for the heater; the
measuring unit is °C.
Example:
Input:
TEMP:60
Result:
The temperature control sets the heating temperature to
60 °C.
Return:
OK
31
TRIGGER:SET:?
Description:
The pulse parameter set currently present in the RAM is displayed in the
following order:
Rising, open time, falling, needle lift, number of pulses, delay.
Parameters relating to time are indicated in 1/10 ms, except for “Falling“
which is given in 1/100 ms. If the valve is currently operated in external
mode, the value for “Open Time“ is “EXTERNAL“. In infinite mode, the num-
ber of pulses is always “0”.
Example:
Input:
TRIGGER:SET:?
Result:
Information is given about the current cycle parameters.
Rising: 10
≙
1,0 ms
(ms = Milliseconds)
Open Time: 10
≙
1,0 ms
Falling: 15
≙
0,15 ms
Needle Lift: 90 %