Veriteq vLog 4.4 User’s Guide
31
1. Under
Clear Mode
, select one of the following options:
Do not clear logger
This is the default setting if the logger memory is
not yet full.
Use this setting when the logger memory is not yet
full and all you want to do is change
Stop
mode set-
tings
without
clearing the existing samples.
If the logger memory is full, or if
Start
mode or
Sample Interval
changes are required, you must
use the
Clear logger
setting to make the necessary
changes.
Clear Logger
This is the default setting if the logger memory is
full.
The
Clear logger
setting allows you to modify
both
Start
and
Stop
mode parameters and change
Sample Interval
values. Choosing this setting re-
sults in the logger’s memory being cleared follow-
ing completion of the setup changes. You will lose
data unless you transfer it first.
If you are in doubt as to whether the information on
the logger is still valuable, first make a copy of the
data using
Logger
>
Transfer
, and then make the
necessary logger setup changes.
When you choose
Clear logger
, the
Start Mode
and
Stop Mode
radio buttons become available.
Summary of Contents for vLog 4.4
Page 1: ...Veriteq vLog 4 4 User s Guide w w w v e r i t e q c o m w w w v a i s a l a c o m ...
Page 6: ......
Page 16: ...Introducing Veriteq vLog 10 ...
Page 17: ...11 2 Installing vLog Software System Requirements 12 Installing vLog Software 12 ...
Page 20: ...Installing vLog Software 14 ...
Page 98: ...Viewing Reports and Statistics 92 ...
Page 99: ...93 7 Troubleshooting Serial Port Problems 94 Frequently Asked Questions 95 ...
Page 103: ...97 8 Glossary 98 A B 98 C E 98 F K 99 L Q 99 R T 99 U Z 101 ...
Page 108: ...Glossary 102 ...