9 Diagnosis, asset management and service
VEGAPULS 67 • Profibus PA
36533-EN-170405
Code
Text message
Reason
Rectification
DevSpec
Diagnosis Bits
M502
Error in the diag-
nostics memory
•
Hardware error EEPROM
•
Exchanging the electronics
•
Send instrument for repair
Bit 15
M503
Measurement reli-
ability too low
•
The echo/noise ratio is too small
for reliable measurement
•
Check installation and process
conditions
•
Clean the antenna
•
Change polarisation direction
•
Use instrument with higher
sensitivity
Bit 16
M504
Error at a device in-
terface
•
Hardware defect
•
Check connections
•
Exchanging the electronics
•
Send instrument for repair
Bit 17
M505
No echo available
•
Level echo can no longer be
detected
•
Clean the antenna
•
Use a more suitable antenna/
sensor
•
Remove possible false echoes
•
Optimize sensor position and
orientation
Bit 21
9.4 Rectify faults
The operator of the system is responsible for taking suitable meas-
ures to rectify faults.
The first measures are:
•
Evaluation of fault messages via the adjustment device
•
Checking the output signal
•
Treatment of measurement errors
Further comprehensive diagnostics options are available with a PC
with PACTware and the suitable DTM. In many cases, the reasons can
be determined in this way and faults rectified.
The below tables show typical examples of application-related
measurement errors with bulk solids. A distinction is made between
measurement errors during:
•
Constant level
•
Filling
•
Emptying
The images in column "
Error pattern
" show the real level as a broken
line and the level displayed by the sensor as a continuous line.
Reaction when malfunc-
tion occurs
Procedure for fault recti-
fication
Treatment of measure-
ment errors with bulk
solids