Vacuum Instrument Corp.
2099 Ninth Avenue
Ronkonkoma, NY 11779
631.737.0900
14
. Reconnect the electrical connector to the source and collector on the mass
spectrometer.
15
. Reconnect the electrical connector to the turbo-molecular pump.
4.8 Calibration
Requirements
Calibration is required when components associated with a measurement
function are serviced, repaired or replaced.
The items that need to be calibrated and reset after service, repair or
replacement are:
•
Test port and fore-line Pirani gauges
• Hi-Vac
gauge
• Temperature
sensor
Test Port Pirani Calibration
NOTE
Calibration is required whenever a gauge is replaced.
1.
Ensure that a blank plug is attached to the test port and that all
valves except the fore-line valve are closed. Verify that the MS-40
is in the Standard mode.
2.
Open the rear panel. To open the rear panel, turn the captive screw
at the top of the panel 1/4 turn counterclockwise and gently lower
the door.
3
.
Using either of the SELECT arrow keys on the left side of the
User Panel, select the Test Port Pressure function. The function
has been selected when the green indicator light to its left is lit.
4
.
Set the unit to manually range down to 10
-10
range (to ensure
that the unit is in direct mode) by pressing the auto/manual
momentary switch so that the manual ranging mode is initiated and
then pressing the “down arrow” key so that the “10 range” is
selected as the eventual “pump down” goal.
5.
Press the START pushbutton to start the test cycle. Allow the unit
to pump down so that it reaches Crossover 2. The fine valve will
be open (the green indicator next to FINE VALVE on the User
Panel will be lit).
1109-186-00 Rev. C
4-19
Summary of Contents for MS40
Page 1: ......
Page 5: ...Figure 1 1 VIC Model MS 40 Portable Automatic Leak Detector 1109 186 00 Rev C 1 2...
Page 9: ...6 3 4 2 5 7 1 9 8 FIGURE 1 2 INTERNAL ROTARY VANE MECHANICAL PUMP 1109 186 00 Rev C 1 6...
Page 17: ......
Page 19: ...FIGURE 2 1 MS 40 USER PANEL 1109 186 00 Rev C 2 2...
Page 32: ...FIGURE 2 2 MS 40 REAR PANEL CONNECTIONS 1109 186 00 Rev C 2 15...
Page 35: ...Figure 2 3 MS 40 Hand Held Remote Control 1109 186 00 Rev C 2 18...
Page 45: ...1109 186 00 Rev C 2 28...
Page 53: ......
Page 60: ...1109 186 00 Rev C 3 7...
Page 69: ......
Page 85: ...1109 186 00 Rev C 4 16...
Page 98: ...1109 186 00 Rev C 4 29...
Page 103: ...1109 186 00 Rev C 4 34...
Page 104: ...1109 186 00 Rev C 4 35...
Page 105: ...1109 186 00 Rev C 4 36...
Page 106: ...1109 186 00 Rev C 4 37...
Page 107: ...1109 186 00 Rev C 4 38...
Page 108: ......
Page 112: ...1109 186 00 Rev C 5 4...
Page 115: ...1109 186 00 Rev C 5 7...
Page 118: ......
Page 122: ......
Page 136: ...1109 186 00 Rev C b ii...
Page 137: ...1109 186 00 Rev C b iii...
Page 138: ......
Page 160: ......
Page 162: ......
Page 164: ...1109 186 00 Rev C f ii...
Page 165: ...1109 186 00 Rev C f iii...
Page 167: ...1109 186 00 Rev C f v...
Page 168: ...1109 186 00 Rev C f vi...
Page 169: ...1109 186 00 Rev C f vii...