NOTE:
Never remove O-rings with a metal tool, as this could
scratch the O-ring groove, and cause a leak. Use a plastic
or wooden (or any soft material) tool to remove the O-ring
by inserting the tool between the inside of the O-ring and
its groove and then sliding the tool around the inside of the
O-ring. This will cause the O-ring to pop up. An O-
ring may have to be held on its opposite side to prevent its
turning in the groove. This procedure usually gives much
better results than trying to pry the O-ring from its groove.
O-rings are susceptible to absorbing large quantities of the solvents used in
cleaning. Subsequent evaporation of these solvents when the system is later
evacuated (during testing) must be avoided, therefore O-rings and gaskets
should be removed from flange grooves and treated as a separate cleaning
problem. O-rings and gaskets are either replaced at the time of re-assembly
(recommended), or carefully wiped clean with a lint-free material and
inspected for surface damage of any kind before they are used again. If a new
O-ring is used, it should be wiped clean, inspected and lubricated with a light
film of vacuum grease, essentially the same procedure as would take place
when reusing an old O-ring. Use vacuum grease as sparingly as possible since
it later may become a source of system contamination.
Cleaning the Mass Spectrometer
Refer to Figure 4-2 for an exploded view of the mass spectrometer.
1
.
Refer to Section 4.9 for an explanation of how to remove the
source from the MS-40.
2
Disassemble the two half sections of the spectrometer housing
and remove the center slit plate and O-ring.
3.
Clean the two sections. The interior of both sections should be
lightly sanded or bead blasted to remove all stains. Blow out the
bead blast residue with dry, oil-free air. The sections should
then be cleaned ultrasonically in freon and dried with a hot air
gun. Make sure to handle the sections when wearing lint free
gloves. Reassemble the two sections and install a new O-ring.
1109-186-00 Rev. C
4-14
Summary of Contents for MS40
Page 1: ......
Page 5: ...Figure 1 1 VIC Model MS 40 Portable Automatic Leak Detector 1109 186 00 Rev C 1 2...
Page 9: ...6 3 4 2 5 7 1 9 8 FIGURE 1 2 INTERNAL ROTARY VANE MECHANICAL PUMP 1109 186 00 Rev C 1 6...
Page 17: ......
Page 19: ...FIGURE 2 1 MS 40 USER PANEL 1109 186 00 Rev C 2 2...
Page 32: ...FIGURE 2 2 MS 40 REAR PANEL CONNECTIONS 1109 186 00 Rev C 2 15...
Page 35: ...Figure 2 3 MS 40 Hand Held Remote Control 1109 186 00 Rev C 2 18...
Page 45: ...1109 186 00 Rev C 2 28...
Page 53: ......
Page 60: ...1109 186 00 Rev C 3 7...
Page 69: ......
Page 85: ...1109 186 00 Rev C 4 16...
Page 98: ...1109 186 00 Rev C 4 29...
Page 103: ...1109 186 00 Rev C 4 34...
Page 104: ...1109 186 00 Rev C 4 35...
Page 105: ...1109 186 00 Rev C 4 36...
Page 106: ...1109 186 00 Rev C 4 37...
Page 107: ...1109 186 00 Rev C 4 38...
Page 108: ......
Page 112: ...1109 186 00 Rev C 5 4...
Page 115: ...1109 186 00 Rev C 5 7...
Page 118: ......
Page 122: ......
Page 136: ...1109 186 00 Rev C b ii...
Page 137: ...1109 186 00 Rev C b iii...
Page 138: ......
Page 160: ......
Page 162: ......
Page 164: ...1109 186 00 Rev C f ii...
Page 165: ...1109 186 00 Rev C f iii...
Page 167: ...1109 186 00 Rev C f v...
Page 168: ...1109 186 00 Rev C f vi...
Page 169: ...1109 186 00 Rev C f vii...