Operator Controls & Indicators
The following Controls & Indicators are located on the User Panel at
the top right front of the unit. Items are listed from left to right, top to
bottom (see Figure 2-1 for an illustration of the User Panel).
LEAK RATE BAR GRAPH
An LED bar graph indicating the mantissa of the measured
leak rate of the test object. For a full leak rate reading, this
meter is used in conjunction with the Numeric Display. The
bar graph display is in standard cubic centimeters per second
or in millibar liters per second.
NUMERIC DISPLAY
A two-digit alphanumeric display with two separate functions.
During testing, the display will normally indicate the exponent
of the current leak rate. For example, if the leak rate is 3 x 10
-7
std cc/sec, the Leak Rate Bar Graph will indicate
3
while the
Numeric Display will indicate the number
7
. The second
function of the display has been programmed to provide error
codes when a system fault has occurred (see Section 4.10 for a
list of error codes, their description and corrective action).
The display will indicate code
EE
while at the same time on
the Service Panel the Error Code indicator will be lit and the
Service Panel alphanumeric display will indicate the error
code.
TEST PORT
A bar graph that displays the pressure at the test port. Scaled
for either millitorr or millibar. The test port pressure is also
displayed in digital format on the Service Panel Alphanumeric
Display.
ACCEPT
During leak testing, this indicator will illuminate when the leak
measured is less than the programmed reject rate (or
maximum acceptable leak rate) and the unit is also in or below
the range of the reject value. The reject rate is set by the
user from the Service Panel (see Section 2.7).
1109-186-00 Rev. C
2-3
Summary of Contents for MS40
Page 1: ......
Page 5: ...Figure 1 1 VIC Model MS 40 Portable Automatic Leak Detector 1109 186 00 Rev C 1 2...
Page 9: ...6 3 4 2 5 7 1 9 8 FIGURE 1 2 INTERNAL ROTARY VANE MECHANICAL PUMP 1109 186 00 Rev C 1 6...
Page 17: ......
Page 19: ...FIGURE 2 1 MS 40 USER PANEL 1109 186 00 Rev C 2 2...
Page 32: ...FIGURE 2 2 MS 40 REAR PANEL CONNECTIONS 1109 186 00 Rev C 2 15...
Page 35: ...Figure 2 3 MS 40 Hand Held Remote Control 1109 186 00 Rev C 2 18...
Page 45: ...1109 186 00 Rev C 2 28...
Page 53: ......
Page 60: ...1109 186 00 Rev C 3 7...
Page 69: ......
Page 85: ...1109 186 00 Rev C 4 16...
Page 98: ...1109 186 00 Rev C 4 29...
Page 103: ...1109 186 00 Rev C 4 34...
Page 104: ...1109 186 00 Rev C 4 35...
Page 105: ...1109 186 00 Rev C 4 36...
Page 106: ...1109 186 00 Rev C 4 37...
Page 107: ...1109 186 00 Rev C 4 38...
Page 108: ......
Page 112: ...1109 186 00 Rev C 5 4...
Page 115: ...1109 186 00 Rev C 5 7...
Page 118: ......
Page 122: ......
Page 136: ...1109 186 00 Rev C b ii...
Page 137: ...1109 186 00 Rev C b iii...
Page 138: ......
Page 160: ......
Page 162: ......
Page 164: ...1109 186 00 Rev C f ii...
Page 165: ...1109 186 00 Rev C f iii...
Page 167: ...1109 186 00 Rev C f v...
Page 168: ...1109 186 00 Rev C f vi...
Page 169: ...1109 186 00 Rev C f vii...