
VENT
Dual function pushbutton. When pressed momentarily, the
MS-40 is interrupted from its current testing cycle and placed
in
standby
(wait) mode. The green READY indicator will
light. When pressed and held for a preprogrammed time (as
programmed by the user; factory default is 0.5 second) the
pushbutton initiates an AUTO-VENT cycle and returns the
test port to atmospheric pressure. The delay may be
programmed for up to 999.9 seconds.
ON/OFF
Turns the MS-40 ON or OFF.
The following Controls & Indicators are accessible from the Service
Panel (the left half of the panel at the top front of the unit). Items are
listed from left to right, top to bottom (see Figure 2-1 for an illustration
of the Service Panel).
SERVICE MODE INDICATOR
When lit, the MS-40 has been placed in Service Mode. The
unit is placed in Service Mode by pressing the Service Mode
Interlock button located on the upper right side of the “A” circuit
board, mounted on the inside of the unit’s rear door.
AMP GAIN
Amplifier Gain. When this function is selected, the
alphanumeric display indicates the setting of the leak rate
amplifier (1.0 to 10.0).
Five different gains are used on the unit. The first,
G1
,
is used for the unit’s gross flow mode; the second,
G2
, is
used for the unit’s reverse flow mode and the third
G3
, is
used for the unit’s direct flow mode.
G4
and
G5
are used
for
sniff
mode.
The five different gain settings can be displayed using
the SELECT momentary switches when the AMP GAIN
function has been selected. Gain (calibration) of
the MS-40 is changed by the SET momentary switches.
Normally G2 and G3 are set automatically by the
MS-40 during a Tune and Calibration cycle, however
Vacuum Instrument Corp.
Vacuum Instrument Corp.
2099 Ninth Avenue
2099 Ninth Avenue
Ronkonkoma, NY 11779
Ronkonkoma, NY 11779
631.737.0900
631.737.0900
the G2 and G3 settings can be manually altered in
Service Mode. The actual setting may be viewed in either
1109-186-00 Rev. C
2-6
Summary of Contents for MS40
Page 1: ......
Page 5: ...Figure 1 1 VIC Model MS 40 Portable Automatic Leak Detector 1109 186 00 Rev C 1 2...
Page 9: ...6 3 4 2 5 7 1 9 8 FIGURE 1 2 INTERNAL ROTARY VANE MECHANICAL PUMP 1109 186 00 Rev C 1 6...
Page 17: ......
Page 19: ...FIGURE 2 1 MS 40 USER PANEL 1109 186 00 Rev C 2 2...
Page 32: ...FIGURE 2 2 MS 40 REAR PANEL CONNECTIONS 1109 186 00 Rev C 2 15...
Page 35: ...Figure 2 3 MS 40 Hand Held Remote Control 1109 186 00 Rev C 2 18...
Page 45: ...1109 186 00 Rev C 2 28...
Page 53: ......
Page 60: ...1109 186 00 Rev C 3 7...
Page 69: ......
Page 85: ...1109 186 00 Rev C 4 16...
Page 98: ...1109 186 00 Rev C 4 29...
Page 103: ...1109 186 00 Rev C 4 34...
Page 104: ...1109 186 00 Rev C 4 35...
Page 105: ...1109 186 00 Rev C 4 36...
Page 106: ...1109 186 00 Rev C 4 37...
Page 107: ...1109 186 00 Rev C 4 38...
Page 108: ......
Page 112: ...1109 186 00 Rev C 5 4...
Page 115: ...1109 186 00 Rev C 5 7...
Page 118: ......
Page 122: ......
Page 136: ...1109 186 00 Rev C b ii...
Page 137: ...1109 186 00 Rev C b iii...
Page 138: ......
Page 160: ......
Page 162: ......
Page 164: ...1109 186 00 Rev C f ii...
Page 165: ...1109 186 00 Rev C f iii...
Page 167: ...1109 186 00 Rev C f v...
Page 168: ...1109 186 00 Rev C f vi...
Page 169: ...1109 186 00 Rev C f vii...