To display
D2
through
D7
, continue pressing the same SELECT
momentary switch; the indicator light will not move to the next
function (in this case, EMISSION) until all seven delay
parameters have been cycled. When the desired delay
parameter appears in the alphanumeric display, it can be
adjusted as in
Step 3.
2.8 Setting Additional Parameters
Amplifier Gain
When this function is selected, the alphanumeric display indicates the gain
setting of the leak rate amplifier (from 0.1 to 10.0). The amplifier gain is
the multiplier used by the computer to compensate for changes in
sensitivity of the mass spectrometer. The user will only use this function
when calibrating using a known external leak source or when unable to
otherwise properly calibrate the unit.
Five separate gains are available:
G1
refers to the gain while the unit is in
gross flow
mode
G2
refers to the gain while the unit is in
reverse flow
mode
G3
refers to the gain while the unit is in
direct flow
mode.
G4
refers to the gain while the unit is in
sniff
mode.
G5
refers to the gain while the unit is in
power sniff
mode.
To alter the Amplifier Gain settings:
1
While in Service Mode, press either SELECT key on the User
Panel until the indicator light beside AMP GAIN is illuminated.
Continuing
to
press
the
same
SELECT key will display each
gain parameter (G1–G5) in sequence before the indicator
moves to the next function (in this case, ANODE VOLT).
2
When the desired gain parameter is displayed, press either the
left or right SET key to decrease or increase the Amplifier Gain.
NOTE
G1
(Gross Mode gain) may be set in the Standard mode when
the MS-40 is leak testing in gross and sniff mode range.
1109-186-00 Rev. C
2-31
Summary of Contents for MS40
Page 1: ......
Page 5: ...Figure 1 1 VIC Model MS 40 Portable Automatic Leak Detector 1109 186 00 Rev C 1 2...
Page 9: ...6 3 4 2 5 7 1 9 8 FIGURE 1 2 INTERNAL ROTARY VANE MECHANICAL PUMP 1109 186 00 Rev C 1 6...
Page 17: ......
Page 19: ...FIGURE 2 1 MS 40 USER PANEL 1109 186 00 Rev C 2 2...
Page 32: ...FIGURE 2 2 MS 40 REAR PANEL CONNECTIONS 1109 186 00 Rev C 2 15...
Page 35: ...Figure 2 3 MS 40 Hand Held Remote Control 1109 186 00 Rev C 2 18...
Page 45: ...1109 186 00 Rev C 2 28...
Page 53: ......
Page 60: ...1109 186 00 Rev C 3 7...
Page 69: ......
Page 85: ...1109 186 00 Rev C 4 16...
Page 98: ...1109 186 00 Rev C 4 29...
Page 103: ...1109 186 00 Rev C 4 34...
Page 104: ...1109 186 00 Rev C 4 35...
Page 105: ...1109 186 00 Rev C 4 36...
Page 106: ...1109 186 00 Rev C 4 37...
Page 107: ...1109 186 00 Rev C 4 38...
Page 108: ......
Page 112: ...1109 186 00 Rev C 5 4...
Page 115: ...1109 186 00 Rev C 5 7...
Page 118: ......
Page 122: ......
Page 136: ...1109 186 00 Rev C b ii...
Page 137: ...1109 186 00 Rev C b iii...
Page 138: ......
Page 160: ......
Page 162: ......
Page 164: ...1109 186 00 Rev C f ii...
Page 165: ...1109 186 00 Rev C f iii...
Page 167: ...1109 186 00 Rev C f v...
Page 168: ...1109 186 00 Rev C f vi...
Page 169: ...1109 186 00 Rev C f vii...