Cut any tie wraps surrounding the sensor and make sure that the
only object connected to the sensor is the calibrator.
5.
Push the quick release button on the calibrator body. Remove the
temperature sensor and leak standard by pulling the fitting down,
angling it out and then pulling out the fitting while the temperature
sensor is still attached to it.
6.
Remove the temperature sensor from the leak standard and reattach
to a new leak standard.
7.
Install the new leak standard and temperature sensor onto the
calibrator body. Place the fitting in the unit, angling the fitting in
the same manner as in Step 5 and then pressing the fitting into the
calibrator body. A positive click indicates that the standard has
been correctly installed.
Circuit Board Replacement
The MS-40 contains three circuit boards in the unit and one circuit board in the
optional hand held remote. Circuit Board A is mounted on the inside of the
unit’s rear panel, Circuit Board B is mounted on the inside of the unit’s front
panel, Circuit Board C is located underneath and attached to the User Panel
and Circuit Board D is the board inside the handheld remote.
WARNING!
The MS-40 must be turned OFF and disconnected from the
electrical supply prior to any repair or replacement of any
of the circuit boards. When the unit is powered ON,
potentially lethal high voltages are continually applied to
the circuit boards and other areas within the unit during its
operation. Failure to follow this precaution may lead to
severe injury or death.
To replace Circuit Board A:
1.
Open the rear panel by turning the captive screw at the top of the panel 1/4
turn counterclockwise and gently lowering the door.
2.
Loosen and remove the screws that hold the shield that covers the
board. Remove the shield.
1109-186-00 Rev. C
4-31
Summary of Contents for MS40
Page 1: ......
Page 5: ...Figure 1 1 VIC Model MS 40 Portable Automatic Leak Detector 1109 186 00 Rev C 1 2...
Page 9: ...6 3 4 2 5 7 1 9 8 FIGURE 1 2 INTERNAL ROTARY VANE MECHANICAL PUMP 1109 186 00 Rev C 1 6...
Page 17: ......
Page 19: ...FIGURE 2 1 MS 40 USER PANEL 1109 186 00 Rev C 2 2...
Page 32: ...FIGURE 2 2 MS 40 REAR PANEL CONNECTIONS 1109 186 00 Rev C 2 15...
Page 35: ...Figure 2 3 MS 40 Hand Held Remote Control 1109 186 00 Rev C 2 18...
Page 45: ...1109 186 00 Rev C 2 28...
Page 53: ......
Page 60: ...1109 186 00 Rev C 3 7...
Page 69: ......
Page 85: ...1109 186 00 Rev C 4 16...
Page 98: ...1109 186 00 Rev C 4 29...
Page 103: ...1109 186 00 Rev C 4 34...
Page 104: ...1109 186 00 Rev C 4 35...
Page 105: ...1109 186 00 Rev C 4 36...
Page 106: ...1109 186 00 Rev C 4 37...
Page 107: ...1109 186 00 Rev C 4 38...
Page 108: ......
Page 112: ...1109 186 00 Rev C 5 4...
Page 115: ...1109 186 00 Rev C 5 7...
Page 118: ......
Page 122: ......
Page 136: ...1109 186 00 Rev C b ii...
Page 137: ...1109 186 00 Rev C b iii...
Page 138: ......
Page 160: ......
Page 162: ......
Page 164: ...1109 186 00 Rev C f ii...
Page 165: ...1109 186 00 Rev C f iii...
Page 167: ...1109 186 00 Rev C f v...
Page 168: ...1109 186 00 Rev C f vi...
Page 169: ...1109 186 00 Rev C f vii...