20.
Sampling Probe
- A device used in pressure testing and so designed
as to collect tracer gas from a restricted area of the test object and
feed it to the leak detector. Also called
pressure probe
or
sniffer
.
21.
Hood Test
- An overall test in which an object under vacuum test is
enclosed by a “hood” which is filled with tracer gas so as to subject
all parts of the test object to examination at one time.
22.
Isolation Test
- A method of determining whether a leak is present in
a system, or of obtaining an estimate of its magnitude, by observing
the rate of rise of pressure in the evacuated system when the system
is isolated from the pump. Also called
rate of rise
test.
23.
Masking
- The covering of a section of a test object so as to prevent
tracer gas from entering leaks that may exist in the covered section.
24.
Flooded System
- A system that, while being leak tested under
vacuum, becomes so filled with tracer gas as to make impracticable
further
leak
detection
by means of a probe.
25.
Helium Drift
- In the case of leak detection with a helium probe, the
drift of helium to a leak or permeable gasket located at a point
sufficiently remote from the end of the probe to mislead the operator
into suspecting the area near the probe.
26.
Standard Leak
a
. A device that permits leakage through it, at a specified rate, of
a specified gas, with atmospheric pressure at one end of the
device and a pressure on the other side sufficiently low to have
negligible effect on the leak rate.
b
. A capillary or porous wall leak, usually in a glass or metal tube,
whose dimensions have been adjusted to give a conductance
within specified limits for a specified gas at a standard reference
temperature with specified inlet and exit pressures. Standard
leaks for attaching to vacuum test manifolds with air at
atmospheric pressure exposed to the inlet are usually protected
by filters to avoid clogging by dust particles. Standard leaks for
calibrating
mass
spectrometers
are usually fused to a glass
reservoir containing the specified gas at a known high pressure.
c
. A device providing a known throughput into a vacuum system.
Also referred to as a
calibrated leak
.
a-iii
Summary of Contents for MS40
Page 1: ......
Page 5: ...Figure 1 1 VIC Model MS 40 Portable Automatic Leak Detector 1109 186 00 Rev C 1 2...
Page 9: ...6 3 4 2 5 7 1 9 8 FIGURE 1 2 INTERNAL ROTARY VANE MECHANICAL PUMP 1109 186 00 Rev C 1 6...
Page 17: ......
Page 19: ...FIGURE 2 1 MS 40 USER PANEL 1109 186 00 Rev C 2 2...
Page 32: ...FIGURE 2 2 MS 40 REAR PANEL CONNECTIONS 1109 186 00 Rev C 2 15...
Page 35: ...Figure 2 3 MS 40 Hand Held Remote Control 1109 186 00 Rev C 2 18...
Page 45: ...1109 186 00 Rev C 2 28...
Page 53: ......
Page 60: ...1109 186 00 Rev C 3 7...
Page 69: ......
Page 85: ...1109 186 00 Rev C 4 16...
Page 98: ...1109 186 00 Rev C 4 29...
Page 103: ...1109 186 00 Rev C 4 34...
Page 104: ...1109 186 00 Rev C 4 35...
Page 105: ...1109 186 00 Rev C 4 36...
Page 106: ...1109 186 00 Rev C 4 37...
Page 107: ...1109 186 00 Rev C 4 38...
Page 108: ......
Page 112: ...1109 186 00 Rev C 5 4...
Page 115: ...1109 186 00 Rev C 5 7...
Page 118: ......
Page 122: ......
Page 136: ...1109 186 00 Rev C b ii...
Page 137: ...1109 186 00 Rev C b iii...
Page 138: ......
Page 160: ......
Page 162: ......
Page 164: ...1109 186 00 Rev C f ii...
Page 165: ...1109 186 00 Rev C f iii...
Page 167: ...1109 186 00 Rev C f v...
Page 168: ...1109 186 00 Rev C f vi...
Page 169: ...1109 186 00 Rev C f vii...