2.3 Starting the MS-40 Leak Detector
The MS-40 is started from the User Panel simply by pressing the ON/OFF
(I/O) switch. During the three-minute Start-up mode, the following will
occur within the unit:
1
The unit will activate a self-diagnostic routine.
2
The fore-line area will be pumped down.
3
The turbo-molecular pump will spin up to its rated speed.
4
If the filament was on when the unit was last turned off, the
filament will be on again at the end of the start-up mode. If the
filament was off when the unit was turned off, the user will
have to press the FIL ON momentary switch to turn on the
filament.
NOTE:
During the start up procedure, all LED’s on the User Panel will be tested
by being sequentially lit.
When the Start-up mode has been completed, the Automatic Ranging
indicator will be lit, the test port measurement gauge will be activated, the
User Panel numeric display will indicate
4
and the READY light
indicator will be lit.
Pre-Testing Start-Up Procedure:
Ensure that all connections are secure and all utilities are available. Check
that:
•
A.C. power is connected.
•
A helium supply source is ready and available (if
applicable).
•
There is a sufficient supply of oil in both the internal and
external
rough
pumps
•
All external accessories are properly connected.
Vacuum Instrument Corp.
2099 Ninth Avenue
Ronkonkoma, NY 11779
631.737.0900
1109-186-00 Rev. C
2-19
Summary of Contents for MS40
Page 1: ......
Page 5: ...Figure 1 1 VIC Model MS 40 Portable Automatic Leak Detector 1109 186 00 Rev C 1 2...
Page 9: ...6 3 4 2 5 7 1 9 8 FIGURE 1 2 INTERNAL ROTARY VANE MECHANICAL PUMP 1109 186 00 Rev C 1 6...
Page 17: ......
Page 19: ...FIGURE 2 1 MS 40 USER PANEL 1109 186 00 Rev C 2 2...
Page 32: ...FIGURE 2 2 MS 40 REAR PANEL CONNECTIONS 1109 186 00 Rev C 2 15...
Page 35: ...Figure 2 3 MS 40 Hand Held Remote Control 1109 186 00 Rev C 2 18...
Page 45: ...1109 186 00 Rev C 2 28...
Page 53: ......
Page 60: ...1109 186 00 Rev C 3 7...
Page 69: ......
Page 85: ...1109 186 00 Rev C 4 16...
Page 98: ...1109 186 00 Rev C 4 29...
Page 103: ...1109 186 00 Rev C 4 34...
Page 104: ...1109 186 00 Rev C 4 35...
Page 105: ...1109 186 00 Rev C 4 36...
Page 106: ...1109 186 00 Rev C 4 37...
Page 107: ...1109 186 00 Rev C 4 38...
Page 108: ......
Page 112: ...1109 186 00 Rev C 5 4...
Page 115: ...1109 186 00 Rev C 5 7...
Page 118: ......
Page 122: ......
Page 136: ...1109 186 00 Rev C b ii...
Page 137: ...1109 186 00 Rev C b iii...
Page 138: ......
Page 160: ......
Page 162: ......
Page 164: ...1109 186 00 Rev C f ii...
Page 165: ...1109 186 00 Rev C f iii...
Page 167: ...1109 186 00 Rev C f v...
Page 168: ...1109 186 00 Rev C f vi...
Page 169: ...1109 186 00 Rev C f vii...