
Vacuum Instrument Corp.
2099 Ninth Avenue
Ronkonkoma, NY 11779
631.737.0900
1.5.1 Standard Operation
This is the normal operating mode for the MS-40. During operation, the User
need only use the right (uncovered) portion of the display panel (the User
Panel is illustrated in Figure 2-1). All functions to be accessed and display
information needed to leak test an object are found on this panel. The panel
displays bar graphs indicating test port pressure and leak rate values.
Additionally, the panel indicates the current leak range mode, an accept/reject
indicator that indicates if the leak being tested is above or below the test set
point, and whether the unit is in test or standby mode. The user panel also has
toggles for automatic or manual ranging (including greater/lesser
pushbuttons), audio volume pushbuttons, an auto- zero pushbutton, a sniff
mode indicator and START and VENT pushbuttons (START is used to
commence the testing cycle as well as to interrupt the testing cycle without the
chance of accidentally venting the system, while VENT is used to interrupt
testing and/or to vent the test port).
1.5.2 Service Mode
WARNING!
The Service Mode of the MS-40 is intended for use by
qualified Vacuum Instrument personnel or Users specifically
trained by Vacuum Instrument. Under no circumstances
should any individual be assigned access to this mode
without a complete knowledge of vacuum systems as well as
the MS-40 System. Use by an unqualified operator may
result in damage to the unit and/or personal injury.
Service Mode is initiated by first opening the rear panel of the unit (loosen the
1/4-turn screw at the top of the panel, then gently lower the door), and then
pressing the black toggle button at the top right hand side of the circuit board
that is attached to the panel. Once the toggle has been pressed, the unit is
placed in Service Mode and the Service Mode indicator is illuminated.
Service Mode disables the automatic valve controls and allows the User to
manually control all of the valves within the vacuum system. The valves are
controlled by a cluster of labeled momentary switches located on the left of
the User Panel. The LED indicator on each switch is lit when the valve is
open. Included within the vacuum system are: the auto-calibrator valve,
external rough valve, fine (direct) valve, fore-line (reverse) valve, gross testing
valve, purge valve, roughing valve, sniffer valve and the vent valve.
1109-186-00 Rev. C
1-12
Summary of Contents for MS40
Page 1: ......
Page 5: ...Figure 1 1 VIC Model MS 40 Portable Automatic Leak Detector 1109 186 00 Rev C 1 2...
Page 9: ...6 3 4 2 5 7 1 9 8 FIGURE 1 2 INTERNAL ROTARY VANE MECHANICAL PUMP 1109 186 00 Rev C 1 6...
Page 17: ......
Page 19: ...FIGURE 2 1 MS 40 USER PANEL 1109 186 00 Rev C 2 2...
Page 32: ...FIGURE 2 2 MS 40 REAR PANEL CONNECTIONS 1109 186 00 Rev C 2 15...
Page 35: ...Figure 2 3 MS 40 Hand Held Remote Control 1109 186 00 Rev C 2 18...
Page 45: ...1109 186 00 Rev C 2 28...
Page 53: ......
Page 60: ...1109 186 00 Rev C 3 7...
Page 69: ......
Page 85: ...1109 186 00 Rev C 4 16...
Page 98: ...1109 186 00 Rev C 4 29...
Page 103: ...1109 186 00 Rev C 4 34...
Page 104: ...1109 186 00 Rev C 4 35...
Page 105: ...1109 186 00 Rev C 4 36...
Page 106: ...1109 186 00 Rev C 4 37...
Page 107: ...1109 186 00 Rev C 4 38...
Page 108: ......
Page 112: ...1109 186 00 Rev C 5 4...
Page 115: ...1109 186 00 Rev C 5 7...
Page 118: ......
Page 122: ......
Page 136: ...1109 186 00 Rev C b ii...
Page 137: ...1109 186 00 Rev C b iii...
Page 138: ......
Page 160: ......
Page 162: ......
Page 164: ...1109 186 00 Rev C f ii...
Page 165: ...1109 186 00 Rev C f iii...
Page 167: ...1109 186 00 Rev C f v...
Page 168: ...1109 186 00 Rev C f vi...
Page 169: ...1109 186 00 Rev C f vii...