3.11 Pressure control
To prepare valve for PRESSURE CONTROL perform complete
«Setup procedure»
.
The valve has parameters that may be modified to tune pressure control performance.
Refer to
«Tuning of control performance»
.
The included PID controller controls the chamber pressure according to the pressure setpoint by means of the valve
position. The PID controller works with an adaptive algorithm to achieve best results under altering conditions (gasflow, gas
type).
Local operation
:
(‘Control View’ resp. ‘Control Performance Analyzer’)
Remote operation
:
(Refer to chapter «OUTPUT Buffer» for details)
Select or enter pressure setpoint
1. In «OUTPUT Buffer» > «PRESSURE
SETPOINT»
2. Select a valid value
3. In «OUTPUT Buffer» > «CONTROL MODE
SETPOINT»
4. Select [Pressure] (value = 5)