Select from the lists the
Transport
vial number that contains
each
Reagent
fluids (A…D). Five choices are available: None,
Transport 1, Transport 2, Transport 3 and Transport 4.
Enter the number of the
Destination well
to be used as the vial
for adding or removing measured amounts of sample and/or
reagents… The vial number depends on the type of the tray
previously configured in the Galaxie Configuration Manager:
A1 to H12 with 96 well plates.
A1 to P24 with 384 well plates.
A1 to F8 with 48 standard well plates.
When the plate feeder option is installed, it enables to select a
Destination plate
other than the
Sample
plate. In this case, the
available choices are Sample, 1, 2, 3, 4, 5, 6 or 7.
Select from the list if the injection must start at the end of the
injection program (
Start run at end of user program
) or at the
inject marker time defined in the program table (
Start run on
inject marker
).
Use the following buttons to edit the table:
to add a line to the table.
to delete a line in the table.
to clear the entire table.
The following
Actions
can be programmed:
Aspirate from Sample, Air, Destination, Reagent A…D or Wash
a programmed volume (
Amount
) into the buffer tubing. the
maximum
Amount
which can be aspirated is the total volume of
the syringe. The
Speed
of the syringe aspiration can be
selected from 1 to 9, and the syringe
Height
from 0 to 40 mm.