ProStar 430 Autosampler Control Manual
- 45 -
03-914947-33:2
4. The sample loop is quantitatively filled by transporting a
number of times the loop volume through the loop,
depending on the volume of the loop. (3x loop volumes for
loops <=100 µl ; 2x loop volumes for loops from 100 µl to
500 µl ; 1½ x loop volumes for loops > 500 µl).
Full loop injection: Filling the sample loop.
5. The injection valve switches into the Inject position. The
sample loop is now part of the HPLC mobile phase flow
path: sample is transported to the column. The analysis
starts.
Full loop injection: Injecting sample.