ProStar 430 Autosampler Control Manual
- 16 -
03-914947-33:2
Some of the following parameters depend on the injection
mode selected.
The
Flush volume
parameter is the amount of sample used to
fill the needle plus tubing connection to the injection valve (not
the loopfill volume).
Only used for
Full loop
and
Partial
loopfill
injections
.
NOTE:
Flush volumes smaller than twice the volume of the needle plus tubing between
needle and injection valve may result in a decrease of performance.
NOTE:
For more details about the injection modes, refer to the section
Injection Mode
at
the end of this manual.
Wash parameter group:
Select the
Wash mode
in the scrolling list. Three choices are
available: None, After injection (single run) or After sequence
(of injections).
If After injection
is selected, enter the
Wash volume
in the
corresponding field.
If After sequence
is selected, the wash procedure is fixed by
the instrument and the wash volume cannot be set.
Options parameter group:
If a plate cooling has been previously declared in the Galaxie
Configuration Manager, the
Plate cooling
fields can be
activated. Check
Plate cooling
to set the temperature setpoint
(4 to 40°C).
If
Plate cooling automatic stop
is checked, the temperature
control is switched off, then some variations of temperature will
be able to occur during a single run or a sequence.
If this option is not checked, the temperature between the
injections will be kept.