ProStar 430 Autosampler Control Manual
- 50 -
03-914947-33:2
The next injection sequence will start with a flush of 50% of the
programmed flush volume, if more than one injection is to be
made from the same vial and no wash routine is programmed.
Otherwise it will start with a flush of the programmed flush
volume. If the withdrawal of sample for the next injection will
exceed the total volume of the sample buffer tubing, the buffer
tubing is rinsed before the next injection. The next injection will
start with the programmed flush, also see the
Full Loop
Injections
section.
Air segment :
An air segment can be used to reduce the amount of flush
volume. This air segment is at the front of the flush volume and
will not be injected.
Situation after filling the loop in the case of partial loop injections with
(A) or without air segment (B)