ProStar 430 Autosampler Control Manual
- 38 -
03-914947-33:2
¾
Injection valve and ISS unit:
ERROR 11 Injection valve is not in a valid position.
ERROR 12 The injection valve did not switch within 1.5
seconds.
ERROR 13 The switching time of the injection valve exceeds
500 msec.
ERROR 14 ISS valve is not in a valid position.
ERROR 15 The ISS valve did not switch within 1.5 seconds.
¾
Syringe dispenser unit:
ERROR 21 The syringe valve did not switch.
ERROR 22 The syringe did not reach home position in time.
ERROR 23 The syringe spindle did not make the correct
number of rotations.
ERROR 24 The spindle does not rotate.
ERROR 25 The syringe valve did not find a valid position.
¾
Injection needle unit:
ERROR 30 The sample needle arm did not reach or leave
home position (vertical).
ERROR 31 The sample needle arm is in an invalid horizontal
position while moving down.
ERROR 32 The sample needle arm did not reach or leave
destination within a certain time (horizontal).
ERROR 34 Sample needle arm is not in vertical home
position while moving horizontally.
ERROR 39 Vial sensor sticks.