ProStar 430 Autosampler Control Manual
- 22 -
03-914947-33:2
A1 to H12 with 96 well plates.
A1 to P24 with 384 well plates.
A1 to F8 with 48 standard well plates.
NOTE:
If the
Destination well
is used in the mix program, the injection will be performed
from this vial and not from the vial defined in the
Acquisition
sub-method section.
When the plate feeder option is installed, it enables to select a
Destination plate
other than the
Sample
plate. In this case, the
available choices are Sample, 1, 2, 3, 4, 5, 6 or 7.
Check
Mix first injection only
to perform the mix program only
before the first injection from the vial. When unchecked, the mix
program will be performed before each injection.
Use the following buttons to edit the table:
to add a line to the table.
to delete a line in the table.
to clear the entire table.
The following
Actions
can be programmed:
Aspirate from Sample, Air, Destination or Reagent A…D a
programmed volume (
Amount
) into the buffer tubing. The
maximum
Amount
which can be aspirated is the total volume of
the syringe. The
Speed
of the syringe aspiration can be
selected from 1 to 9, and the syringe
Height
from 0 to 40 mm.
Dispense to Sample, Waste, Destination or Reagent A…D a
programmed volume (
Amount
) from the buffer tubing. It is
possible to dispense a larger volume than the volume aspirated
in previous actions. The aspirated
Amount
will be