7. Remove the filter to access the sensors.
1
2
3
4
1
Vaisala HUMICAP
â
sensor. Handle by
the plastic frame.
2
Temperature sensor. Do not touch or
attempt to remove.
3
Sensor socket.
4
Transmitter probe.
8. Pull out the old HUMICAP
â
sensor, and insert the new one.
9. Perform a two-point humidity calibration on the serial line using the
FCRH
command.
10. Insert the new filter.
11. Reassemble, remount, and reconnect the transmitter.
12. Power up the transmitter.
More Information
‣
Two-Point Humidity Calibration and Adjustment Using Computer and HMK15 Humidity
7.9 Filter Selection
When replacing the filter, note that there are two alternative filter options depending on the
probe model:
• HMS and HMD products use the membrane filter (Vaisala order code ASM210856SP)
• HMW products use the porous PTFE filter (Vaisala order code DRW239993SP)
HMDW110 User Guide
M211726EN-D
68