Modifier
Description
cs4
Modulus-65536 checksum of message sent so far, ASCII encoded hexadecimal
notation
csx
NMEA xor-checksum of message sent so far, ASCII encoded hexadecimal
notation
You can also use the hash character # instead of the backslash character \.
6.6.3 Serial Line Communication
Table 20 Open Command
Syntax
Description
open [aaa]
<cr>
Open a connection to a device at the specified
address. Required when device is in poll mode.
Address range 0 ... 255.
Example:
open 5
HMDW110 5 line opened for operator commands
Table 21 Close Command
Syntax
Description
close
<cr>
Close the connection that was opened with the
open command.
Example:
close
line closed
Table 22 Addr Command
Syntax
Description
addr
<cr>
Show current device address and prompt for a
new address.
addr [aaa]
<cr>
Set new device address.
aaa = 0 ... 255.
Chapter 6 – Using Service Port With Terminal Application
43
Summary of Contents for hmdw110 series
Page 1: ...M211726EN D User Guide Vaisala Humidity and Temperature Transmitter Series HMDW110 ...
Page 5: ...Recycling 87 Table of Contents 3 ...
Page 87: ...Appendix A Modbus Reference 85 ...
Page 88: ...HMDW110 User Guide M211726EN D 86 ...
Page 90: ...HMDW110 User Guide M211726EN D 88 ...
Page 91: ......
Page 92: ...www vaisala com ...