Syntax
Description
Example (shows default state without user adjustments - offset is 0 and gain is 1 for both humidity
(Cp offset and gain) and temperature):
l
Cp offset : 0.0000
Cp gain : 1.0000
T offset : 0.0000
T gain : 1.0000
Table 38 Li Command
Syntax
Description
li
<cr>
Enter values for offset and gain parameters for
user adjustment. Useful for restoring some
earlier state of user adjustment.
Use this command only to restore values you
have previously written down based on the
output from the l command, or to restore the
default offset and gain.
Example (shows adjustment has been applied to humidity measurement offset, overwrites it with
0):
li
Cp offset : -1.50922060E-01 ?
0
Cp gain : 1.0000 ?
T offset : 0.0000 ?
T gain : 1.0000 ?
Chapter 6 – Using Service Port With Terminal Application