9. Give the
crh
command to start the calibration and adjustment sequence. The transmitter
shows the measured RH value and prompts you to enter the real humidity of the first
reference point.
crh
RH : 11.5378 1. ref ?
10. Press
Enter
to refresh the measured value (to see if the measurement is now stable).
When it is, enter the value of the first reference point and press
Enter
.
RH : 11.5378 1. ref ?
<cr>
RH : 11.5379 1. ref ?
<cr>
RH : 11.5379 1. ref ?
11.3
Press any key when ready ...
To exit the procedure without adjusting the measurement, press the
Esc
key
on your keyboard.
11. Transfer the probe to the NaCl (75 %RH) salt jar. Plug the hole on the LiCl (11 %RH) jar.
12. Wait for the humidity to stabilize. Press
Enter
on the keyboard to refresh the measured
value:
Press any key when ready ...
RH : 74.6984 2. ref ?
<cr>
RH : 74.7173 2. ref ?
<cr>
13. When the measurement is stable, enter the value of the second reference point and press
Enter
.
RH : 74.9684 2. ref ?
75.4
OK
14. If the adjustment is successful, the transmitter responds with the text
OK
, and stores the
adjustment.
If the reference points do not fulfill the requirements, or the measurement is too unstable
for adjustment, the adjustment sequence will exit with an error message. In such a case,
the adjustment is not stored.
Chapter 7 – Maintenance
61
Summary of Contents for hmdw110 series
Page 1: ...M211726EN D User Guide Vaisala Humidity and Temperature Transmitter Series HMDW110 ...
Page 5: ...Recycling 87 Table of Contents 3 ...
Page 87: ...Appendix A Modbus Reference 85 ...
Page 88: ...HMDW110 User Guide M211726EN D 86 ...
Page 90: ...HMDW110 User Guide M211726EN D 88 ...
Page 91: ......
Page 92: ...www vaisala com ...