Table 32 Crhclr Command
Syntax
Description
crhclr
<cr>
Clear the current user adjustment for humidity.
Factory calibration remains.
Example:
crhclr
OK
Table 33 Ct Command
Syntax
Description
ct
<cr>
Start the two-point temperature calibration and
adjustment sequence. For a full adjustment
procedure, see
Calibration and Adjustment Using Computer
(page 62)
.
ct [ref]
<cr>
Perform a one-point adjustment at the current
temperature. The device calculates the required
correction based on the reference temperature
that you enter.
ref = the correct temperature (in degrees
Celcius) at the environment where the sensor is
now.
Example :
ct
T : 22.9424 1. ref ?
23.0
Press any key when ready ...
T : 54.9873 2. ref ?
55
OK
Table 34 Ctclr Command
Syntax
Description
ctclr
<cr>
Clear the current user adjustment for
temperature. Factory calibration remains.
Example:
ctclr
OK