15. Give the
l
command to verify that the values of the adjustment parameters
Cp offset
and
Cp gain
have changed.
Example: User adjustment parameters after two-point humidity adjustment (your values
will be different):
Cp offset : -2.21914053E-03
Cp gain : 1.0000
T offset : 0.0000
T gain : 1.0000
16. Remove the probe from the salt jar and reconnect the filter. Plug the hole on the salt jar.
17. Update the calibration date and calibration information on the transmitter using the
cdate
and
ctext
commands.
Example: Set calibration date to 2018-07-04 and calibration info text to "Lab2/Mike".
cdate 20180704
Cal. date : 20180704
ctext Lab2/Mike
Cal. info : Lab2/Mike
18. Disconnect the connection cable from the service port.
19. Reconnect the terminal blocks and close the cover.
7.6 Two-Point Temperature Calibration and
Adjustment Using Computer
You can also use Vaisala Insight software for performing calibration and
adjustment. See
Vaisala Insight Software (page 29)
HMDW110 User Guide
M211726EN-D
62