6
SYSTEM SETTINGS
128
Chapter
6
SY
STE
M
SETTINGS
⑥
FLT. Check
Check the filter circuit.
- Analog Filter
Voltage/current input (Load input):
When the input is changed from 0V to 10V or
0mA to 10mA (step input), use the following alues as a guide for normal.
30Hz:
23±2
100Hz:
7±2
300Hz:
3±1
1kHz:
1±1
⑦
PUI Check
Check that the encoder input is recognized.
Operation
Main screen
→
Setting
→
First Setting
→
System Setting
→
Self-Check
→
PUI Check
⑧
COM. Check
Check that RS-232C reception and transmission are in proper
operation.
Press the Trns. Key, the same statement as in display read is
transmitted, and the received data is displayed in the Rx Data
section. Also, in the case of a parity error/framing error, the Parity/
Frame lamp turns red.
Operation
Main screen
→
Setting
→
First Setting
→
System Setting
→
Self-Check
→
COM. Check
⑨
SDC Check (Only when the SD card slot option is mounted)
Check the operation of the SD memory card.
Operation
Selectable range (Start, Err Clear)
Main screen
→
Setting
→
First Setting
→
System Setting
→
Self-Check
→
SDC Check