4
SETTINGS AND CALIBRATIONS
51
Chapter
4
SETTINGS AND CALIBRATIONS
3.
X-axis Start Point
Set the X-axis start point of the waveform.
The setting range differs according to the setting of Wave
Reference and Measurement Length.
Setting range
Time or displacement (Front)
(0 to 2000 × Measurement Length/2000)
Operation
Displacement (Back)
(-2000 to 0 × Measurement Length/2000)
Main screen
→
Setting
→
Work Setting
→
Display Setting
→
X Start Point
4.
X-axis End Point
Set the X-axis end point of the waveform.
The setting range differs according to the setting of Wave
Reference and Measurement Length.
Setting range
Time or displacement (Front)
(X-axis start point + 25, 50, 100, 200, 400, 600, 800, 1000, 1200, 1400, 1600,
1800, 2000, 2200 × Measurement Length/2000)
Displacement (Back)
(X-axis start point + -25, -50, -100, -200, -400, -600, -800, -1000, -1200, -1400,
Operation
-1600, -1800, -2000, -2200 × Measurement Length/2000)
Main screen
→
Setting
→
Work Setting
→
Display Setting
→
X End Point
Setting from Waveform Check
The waveform display axis can also be adjusted by the following operation.
Operation
Main screen
→
Check
→
Display