When the X-axis represents time
2
Measurement Length
Set the time for waveform acquisition.
Operation
Setting range (0.5, 1.0, 2.0, 5.0, 10.0 s)
Main screen
→
Setting
→
First Setting
→
X-axis Setting
→
Measure Length
When the X-axis represents displacement
2
Sensor Phase Select
Select the output phase of the rotary encoder.
Operation
Setting range (A/B-Phase, Only A-Phase)
Main screen
→
Setting
→
First Setting
→
X-axis Setting
→
Phase Select
- Only A-Phase:
A-Phase of the sensor is used. Resolution is 1 pulse.
(Positive counting only.)
- A/B-Phase:
A/B-Phase of the sensor is used. Resolution is 1/4 pulses.
The counting conditions are as follows.
< In case of time >
Time
Load
< In case of displacement >
Displacement
Load
Time
A-Phase OFF
ON
Up-counting
(When the A-phase advances)
1 pulse
B-Phase OFF
ON
A-Phase OFF
ON
Down-counting
(When the B-phase advances)
B-Phase OFF
ON
1/4 pulses