Wavelength Validity Test
No WL(nm) Peak(nm) %T Result
1 241.4
2 361.0
3 417.0
4 537.6
5 641.4
6 807.4
Fig 97
3. Press
【
F3
】
to set tolerance (Fig 98). Input the value by pressing the
numeric keypad.
D2
W
Cell #1
Wavelength Validity Test
WL : 900.0nm %T:
12: 35: 27
Input tolerance:0.8
No WL(nm) Peak(nm) %T Result
1 241.4
2 361.0
3 417.0
4 537.6
5 641.4
6 807.4
Select
Fig 98
4.
Press 【
0Abs/100%T
】for b
lanking.
5. Put the sample (calibrated holmium liquid) into the light path. Press
【
START
】
to check. The results will be displayed on the screen (Fig
99). If the discrepancy between the results and the calibrated values
is not more than the tolerance, “pass” will be displayed after the test
results. Otherwise, “fail” will be displayed.
D2
W
Cell #1
WL : 900.0nm %T:
12: 35: 27
F1:
Standard
F2:
Mode
F3:
Tolerance
Wavelength Validity Test
No WL(nm) Peak(nm) %T Result
1 241.4
239.7
40.06 pass
2 361.0 360.4
42.82 pass
3 417.0
416.9
37.63 pass
4 537.6
537.2
16.50 pass
5 641.4
641.3
24.33 pass
6 807.4
807.7
92.38 pass
Fig 99
6.The result can be saved,loaded and printed by pressing
【