14
1. Abs mode
Push the blank cuvette into the light path.Press
【
F2
】
to
select Abs
mode ,Press
【
0Abs/100%T
】
for Blanking , and then Push the sample into
lightpath to take reading(Fig 20)
2. T% mode
The operation is the same as Abs test mode but pressing
【
F2
】
to
select
T% mode .
3. Conc/Factor mode
Press
【
F1
】
to select a concentration unit (Fig 21). If no unit is
suitable for your test, please select the item “Other”, press enter and input a
new unit by pressing the numeric keypad (Fig 22).
WL : 656.1nm
12: 35: 27
D2
W
Cell #1
Max E
Please select unit:
%
0.000 mg/ml
F factor 2.000
Fig 21
WL : 656.1nm
12: 35: 27
D2
W
Cell #1
Max E
Please input self defined unit:
c
0.000 mg/ml
F factor 2.000
Fig 22
4.
Push the blank cuvette into the light path and press
【
0Abs/100%T
】
for
Blanking. There are now two choices for you to take:
4.1 Press
【
F3
】
to
input known F value, Fig 23. Then push the sample into
lightpath to take reading of concentration
4.2 Push sample of known concentration into the lightpath
Press
【
F4
】
to
input known Conc value, Fig 24. Then push the sample
into lightpath to take reading of concentration.
Note:1.You can select wavelength at any time by pressing
【
SET
λ
】.
After your selection, instrument always blanks automatically.
2.If F value is more than 9999,the “out of range” will display on
screen.