16
J
How to operation
1. Press
【
F1
】
to select unit of concentration (Fig 27).
WL : 700.0nm Abs:
12: 35: 27
D2
W
Cell #1
Please select unit: g/L
Quantitative Test
ID Abs Conc. (mg/L)
WL(nm)
700.0
C=1.000*A^1
r=1.000
Search
Scroll
Fig 27
2 Press
【
SET
λ
】
to select correction methods and enter the
wavelength. There are three correction methods (single,
Isoabsorbance and 3 point, Fig 28) .
Note: Please refer to the Appendix C for the correction method.
WL : 700.0nm Abs:
12: 35: 27
D2
W
Cell #1
Quantitative Test
ID Abs Conc. (mg/L)
WL(nm)
700.0
C=1.000*A^1
r=1.000
Correction method:
Single WL
Search
Scroll
Fig 28
3.Press
【
F2
】
in Fig 26 for more items to select .See Fig 29.
WL : 700.0nm Abs:
12: 35: 27
D2
W
Cell #1
Calibration table
No Conc. (mg/L) Abs
WL(nm)
700.0
C=1.000*A^1
r=1.000
F 1
:Method
F2:
Params
F3:
Standard
F4:
Curve
Fig 29
3.1 Press
【
F1
】
in Fig 29 to select fitting method. There are 4 methods for
you to choose: Linear fit, linear fit through zero, square fit and cubic fit.
3.2 Press
【
F2
】
in Fig 29 to enter directly a known standard curve.Fig29A.