27
D2
W
Cell #1
Tim : 180s Abs:
12: 35: 27
0 Time(s) 180
Total T
180s
Inteval
1.0s
0
A
b
s
1
.0
00
XScale
YScale
F1
:
Setup
F2
:
Mode
F3
:
Process
F4
:
Search
Fig 56
5. Press
【
F3
】
to process the data, and e nter “Begin Time”, ”End Time”
and ”Factor” (Fig 57) and the value in I.U. will be calculated and
displayed (Fig 58). The average straight line between the Begin Time
and End Time will be calculated. The gradient of this line gives the rate
of change of ?A/min.
Note: I.U.=Factor
×Δ
A/min
D2
W
Cell #1
Tim : 60s Abs:
12: 35: 27
0 Time(s) 180
Total T
180s
Inteval
1.0s
0
A
b
s
3
.0
00
XScale
YScale
Begin Time:0
Fig 57
D2
W
Cell #1
Tim : 180s Abs:
12: 35: 27
0 Time(s) 180
Total T
180s
Inteval
1.0s
I.U.=
+0.364
0
A
b
s
1
.0
00
XScale
YScale
F1
:
Setup
F2
:
Mode
F3
:
Process
F4
:
Search
Fig 58
6. If you want to change the scale, please refer to step 5 of “WL scan”.
7. Press
【
F4
】
to search the Abs/%T value in relation to the time axis.
Search point to point by pressing the key
【
<
】
or
【
>
】
. Please refer
to step 6 of “WL scan”.