5
Packing List
Description Quantity
Ÿ
Spectrophotometer.........................................................
1
Ÿ
Mains Lead......................................................................
1
Ÿ
Cuvettes...........................................................................
Set of 4, glass
....................................................................
Set of 2, quartz
Ÿ
Dust Cover.......................................................................
1
Ÿ
Manual .............................................................................
1
Specifications:
Ÿ
Wavelength Range:
190-1100nm
Ÿ
Spectral Bandpass:
4nm
Ÿ
Wavelength Accuracy:
±
0.8nm
Ÿ
Wavelength Repeatability:
±0.05nm
Ÿ
Stray Radiant Energy:
<
0.15%@220nm&340nm
Ÿ
Photometric Range:
0-200%T,-0.3-.2.8A
Ÿ
Noise: <0.001A @ 500nm 0A
Ÿ
Drift:
<0.002A/h @ 500nm
Ÿ
Power Requirements:
AC 110V/60Hz or 220V/50Hz
Ÿ
Dimensions:
550W
×
420L
×
270H
Ÿ
Light Source:
Tungsten Halogen/Deuterium
Ÿ
Weight:
18kg
Installation:
1. After carefully unpacking the contents, check the materials with the
packing list (page 4) to ensure that you have received everything in
good condition.
2. Place the instrument in a suitable location away from direct sunlight.
In order to have the best performance from your instrument, keep it as
far as possible from any strong magnetic or electrical fields or any
electrical device that may generate high-frequency fields. Set the unit
up in an area that is free of dust, corrosive gases and strong vibrations.
3. Remove any obstructions or materials that could hinder the flow of air
under and around the instrument.
4. Use the appropriate power cord and plug into a grounded outlet.
5. Turn on your UNICOUV-2802 model spectrophotometer. Allow it to
warm up for 15 minutes before taking any readings.We suggest you
then do the Calibrate System with the Search 656.1nm to set the
wavelength to the deuterium lamp emission line.
NOTE:
This symbol means Caution,Risk of Danger.Refer to this
Manual(see Appendix B – Lamp Replacement)