UPO2000CS series User Manual
76
(1) Function Introduction
Press UTILITY, then press F5 to enter the secondary menu, then press F2 to enter the test menu.
Pass Test Menu
Functions Options
Descriptions
Allow Test
Close
Close Pass Test
Open
Open Pass Test
Output
Fail
Set the Pass/Fail interface on the rear panel to a "failure" pulse
and produce a buzzer
Pass
Set the Pass/Fail interface on the rear panel to a "pass" pulse
and produce a buzzer
Source
CH1
、
CH2
、
CH3
、
CH4
Set the source of the pass test
Display
Info
Close
Turn off display information
Open
Turn on display information
1/2
Enter Pass Test Menu (2)
Pass Test Menu (2)
Functions Options
Descriptions
Operation
Close
Disable Pass Test
Open
Enable Pass Test
Stop
Setting
Enters Stop Settings Menu
Template
Setting
Enters Template Settings Menu
2/2
Enter Pass Test Main Menu
After opening the pass test, you need to set the stop settings and template settings before you can perform the test
function. Otherwise you'll be prompted "Function is Disabled". The stop setting and template settings menus are
shown below.
Stop Settings Menu
Functions Options
Descriptions
Stop Type
Pass Count
The test function stops after it reaches a specified pass count
number
Fail Count
The test function stops after it reaches a specified fail count
number
Condition
>=
、
<=
Stop condition setting
Summary of Contents for UPO2000CS Series
Page 1: ...1 ...
Page 64: ...UPO2000CS series User Manua 63 X Y Phase Difference Table ...
Page 69: ...UPO2000CS series User Manual 68 User Defined Parameter Statistics ...
Page 92: ...UPO2000CS series User Manua 91 This user manual may be revised without prior notice ...
Page 93: ...UPO2000CS series User Manual 92 ...
Page 94: ...UPO2000CS series User Manua 1 ...