UPO2000CS series User Manual
36
Slope
Rising
Triggers when at the rising edge, signal level meets the set level
Falling
Triggers when at the falling edge, signal level meets the set level
Free Time 8ns~10s
Set the free time internal
Next Page
Go to second page of settings
N-edge Trigger Menu (2)
Functions Options
Descriptions
N
1~65535
Set N value to be between 1~65535
Trigger
Setting
Enters trigger setting menu
Return
Return to N-edge trigger main menu
3.13 Code Pattern Trigger
Code pattern trigger occurs when the specified pattern is identified by the specified code type. The code type is the
combination of the channel logic "AND", each channel can be set to H (high), L (low), X (ignored). You can also
specify a path in the code type for a rising edge or falling edge (only one edge can be specified). If the code pattern
of the other channels are "true" (i.e., the actual code is consistent with the default code type), the oscilloscope will
trigger on the specified edge. If the edge is not specified, the oscilloscope will be triggered at the last edge of the
code type "true". If all channels are set to "ignore", the oscilloscope will not trigger.
Summary of Contents for UPO2000CS Series
Page 1: ...1 ...
Page 64: ...UPO2000CS series User Manua 63 X Y Phase Difference Table ...
Page 69: ...UPO2000CS series User Manual 68 User Defined Parameter Statistics ...
Page 92: ...UPO2000CS series User Manua 91 This user manual may be revised without prior notice ...
Page 93: ...UPO2000CS series User Manual 92 ...
Page 94: ...UPO2000CS series User Manua 1 ...