S-2
ULVAC CRYOGENICS INCORPORATED
Ensure to discharge helium gas and remove either side of the self-sealing couplings from
the used adsorber before disposing of it.
When the refrigerator system is not in operation, the helium gas pressure inside the
adsorber is as high as the fill pressure of the compressor. Disposing the adsorber with
helium gas inside may cause accidents. For example, if the adsorber was thrown in a
burner reactor without discharging helium gas by mistake, the inside pressure of the
adsorber would rise as the inside temperature rises and it may explode. If the adsorber
was pressed with helium gas still inside, it may also cause an explosion. Refer to
“Disposal Consideration” for more information.
Use appropriate charging adaptor to remove helium gas safely.
☆
We collect used adsorbers. If there is any used absorber, please contact a person in your
local dealership or Ulvac Techno Ltd. to have it picked up or send it directory to our
Service Engineering Division. We only accept CTI adsorbers and our adsorbers.
・
Power supply voltage must be kept within the allowable range(refer to
“Section 1 Table1-1
Compressor Unit Standard Specifications
” . Applying more than the allowed voltage
may damage the equipment.
・
Install a ground-fault circuit interrupter(rated current: 10A, rated tripping leakage
current:30mA 3
φ
, rated voltage: 200V).
・
Grounding conductor (earth wire) must be connected (D class grounding (with the
ground resistance of
100Ω)).
・
Connect the refrigerator cable and remote cables (signal lines) to the compressor before
connecting the compressor to the power supply. These wiring connections must not be
done when the compressor is powered on. It could lead to a serious accident such as
electric shock and/or damage to the equipment.
・
Follow recommended sizes for wires and cables.
・
Power lines and signal lines must be wired at least 10cm separate from one another.
5.
福 福
6.
絈 絈
7.
福 福
8.
絈 絈
4.
Disposal of used adsorber
1.
福 福
2.
絈 絈
3.
福 福
4.
絈 絈
5.
Electrical wiring connections
Summary of Contents for C10T
Page 2: ......
Page 8: ...F 6 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Page 14: ...S 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Page 16: ...IW 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Page 32: ...Operation 4 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Page 44: ...Troubleshooting A 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Page 47: ...Compressor Unit Instruction Manual ULVAC CRYOGENICS INCORPORATED B 3 B 4 ...
Page 48: ...Compressor Unit Instruction Manual ULVAC CRYOGENICS INCORPORATED B 5 B 6 ...
Page 50: ...Flexible Hose C 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Page 54: ...CS 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Page 56: ...This page intentionally left blank R 2 ULVAC CRYOGENICS INCORPORATED ...