Installation
0
1
2
3
4
5
6
0.10
(
*
) The water pressure drop (
Δ
P) means differential pressure between cooling water
inlet pressure and outlet pressure.
Figure 3-2
Cooling Water Requirement
A dashed arrow (
→
) in the graph shows the minimum required water flow rate
(3.5L/min.) and water pressure drop (0.04MPa) when the temperature of the
water
EXAMPLE
Press
ure drop
Δ
P
(*
)
(
MP
a
)
0.08
0.06
0
0.072
0.02
0.04
0
5
10
15
20
30
32
25
Flow rate
(
L
/
min
)
35
0 1 2
3
4
5
6
Allowable operating range
Te
mp
erature
(
°C
)
3.5
3-4