Compressor Unit Instruction Manual
ULVAC CRYOGENICS INCORPORATED
4-1
4. OPERATION
4.1 Before Operation
Before operating the compressor, check and verify that:
1. Input power requirements are within the specified values shown in Table 1-1.
2. STOP switch (orange) lights up when the power is supplied to the compressor.
Please be noted that the switch will light up (dark light) even the phase is reversed. If the
compressor does not start when the START switch has been pressed, there is possibility of
reversed phase. Change any one set of RST phases.
3. REMOTE/LOCAL switch is properly set.
4. Cooling water requirements is within the specified values.
5. Helium gas static pressure is within the specified values.
4.2 Normal Operation
Verify that the operating pressure is within the specified values shown in Table 1-1.
CAUTION
Summary of Contents for C10T
Page 2: ......
Page 8: ...F 6 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Page 14: ...S 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Page 16: ...IW 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Page 32: ...Operation 4 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Page 44: ...Troubleshooting A 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Page 47: ...Compressor Unit Instruction Manual ULVAC CRYOGENICS INCORPORATED B 3 B 4 ...
Page 48: ...Compressor Unit Instruction Manual ULVAC CRYOGENICS INCORPORATED B 5 B 6 ...
Page 50: ...Flexible Hose C 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Page 54: ...CS 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Page 56: ...This page intentionally left blank R 2 ULVAC CRYOGENICS INCORPORATED ...