SonixTouch Q+ Service Manual
00.053.204, Revision A
Chapter 7: Field Service Components
7-1
CHAPTER 7: FIELD SERVICE COMPONENTS
This section describes how to handle all field serviceable components, including the modulo, LCD display, console, speakers, UPS,
battery and transducers.
7.1
PROTECTING PATIENT DATA
Before doing any work with the modulo, be sure to complete sections
) in order to the export/backup
and therefore protect the integrity of each client's data.
Once all work has been completed, refer to sections
(or
) to import/restore the client's settings and data.
7.1.1
Exporting User Data (As Required)
Before servicing the modulo, export
User Data
to an external USB backup medium (memory stick, external hard
drive, etc.). This will ensure access to this data if either the modulo or hard drive needs replaced during servicing.
The following options are available for export:
• Imaging Presets
• 3D/4D Presets
• DICOM Server Configuration
• Settings
• Measurement Order
• Obstetrical Tables
• Protocols
• General
• Preset Assignments
• Measurement Customization
• Worksheets
• Banners
•
System Logs
•
Licenses
(
license.key
file).
Equipment/Tools Required:
• USB media (memory stick, external hard drive, etc.).
Note:
Use only one method to
Export/Import Patient/Exam
data:
• when exporting with
, import with
, restore with
.
Note:
When creating a backup prior to servicing or replacing the modulo, Ultrasonix recommends selecting all
options for
Export
.
Note:
Standard
Worksheets
are only available for the
EMED
,
Anesthesia
and
Endocrinology
Protocols
.
SonixHub Worksheets
are available for a wider variety of
Protocols
.
All
Worksheets
are exported when this option is selected.
Note:
System Logs
cannot be imported.
Summary of Contents for SonixTouch Q+
Page 1: ...Analogic Ultrasound Service Manual SonixTouch Q Ultrasound System...
Page 2: ......
Page 4: ......
Page 14: ...Chapter 2 System Specifications 00 053 204 Revision A SonixTouch Q Service Manual 2 2...
Page 42: ...Chapter 4 Performance Testing 00 053 204 Revision A SonixTouch Q Service Manual 4 4...
Page 120: ...Chapter 8 DICOM 00 053 204 Revision A SonixTouch Q Service Manual 8 24...
Page 147: ......