Chapter 11: Maintenance
00.053.204, Revision A
SonixTouch Q+ Service Manual
11-6
To Clean the System Filter:
1.
Power off and unplug the system.
2.
Taking care not to rip/damage the filter, g
ently
pull on the two aluminum system filter frame handles until the
filter is free of its slot on the right side of the system.
3.
Vacuum thoroughly and
gently
reinstall the filter.
4.
Plug in and power on the system.
11.2.11.2Cleaning the Modulo Fan and Exhaust Filters
Equipment/Tools Required:
• #2 Phillips screwdriver.
To Clean the Modulo Fan and Exhaust Filters:
1.
Power off and unplug the system.
2.
Remove the side shrouds (
) and the modulo (
).
3.
After removing the modulo, use the #2 Phillips screwdriver to remove the three exhaust fan filters from the side
of the modulo.
4.
Vacuum both sides, then reinstall the filters.
Caution:
When reinstalling, ensure the filter side faces toward the front of the system and the grid covering
the back of the filter faces towards the back.
Modulo Exhaust Filters
Summary of Contents for SonixTouch Q+
Page 1: ...Analogic Ultrasound Service Manual SonixTouch Q Ultrasound System...
Page 2: ......
Page 4: ......
Page 14: ...Chapter 2 System Specifications 00 053 204 Revision A SonixTouch Q Service Manual 2 2...
Page 42: ...Chapter 4 Performance Testing 00 053 204 Revision A SonixTouch Q Service Manual 4 4...
Page 120: ...Chapter 8 DICOM 00 053 204 Revision A SonixTouch Q Service Manual 8 24...
Page 147: ......