Chapter 3: System Installation
3-13
3.3.2
EMI Filter and Power Panel
The system is delivered with a voltage rating label over the power switch and EMI filter. The precise wording on the
label is determined by the voltage setting on the EMI filter.
Double check that the voltage rating on the label matches the voltage configuration on the EMI filter, then remove
the label.
Figure 3-3: Voltage Rating Label
3.3.2.1 Voltage Configuration
shows the location of the EMI Filter which houses the system's fuses and the main power switch. The
EMI Filter also controls the system voltage setting.
When changing the voltage settings, the EMI filter must be adjusted to the relevant setting and the fuses must be flipped
to match the selected voltage. Refer to
for details on how to remove the fuse box and replace the fuse(s).
Figure 3-4: EMI Filter
Figure 3-5: EMI Filter and System Power Button Location
Table 3-2: System Power Panel
Caution:
The system voltage setting is configured in the factory. It should not need to be changed in the field, but
it should be double-checked.
If the settings are incorrect, refer to the
SonixTouch Q+ Service Manual
for details on voltage and fuse
configuration (
Caution:
The system voltage setting is configured at the factory. It should not need to be changed in the field.
1
Main Power Switch
2
Peripheral Receptacle (AC Power Cord)
1
2
Summary of Contents for SonixTouch Q+
Page 1: ...Analogic Ultrasound Service Manual SonixTouch Q Ultrasound System...
Page 2: ......
Page 4: ......
Page 14: ...Chapter 2 System Specifications 00 053 204 Revision A SonixTouch Q Service Manual 2 2...
Page 42: ...Chapter 4 Performance Testing 00 053 204 Revision A SonixTouch Q Service Manual 4 4...
Page 120: ...Chapter 8 DICOM 00 053 204 Revision A SonixTouch Q Service Manual 8 24...
Page 147: ......