SonixTouch Q+ Service Manual
00.053.204, Revision A
Chapter 3: System Installation
3-1
CHAPTER 3: SYSTEM INSTALLATION
3.1
INSTALLATION ISSUES AND REQUIREMENTS
3.1.1
Environmental Requirements
Verify the system is to be operated in a room that meets the environmental requirements listed in
Table 3-1: Environmental Specifications
Table 3-2: Barcode Reader
Table 3-3: Wireless
Caution:
Avoid placing the system against walls or structures that may decrease air circulation.
Operational Temperature
50º to 104º F (10º to 40º C)
Operational Humidity
30 to 75% relative humidity
Operational Pressure (kilopascal)
70 kPa to 106 kPa
Shipping/Storage Conditions
+5° to +122° F (-15° to +50° C)
Shipping/Storage Humidity
10% to 90% (non-condensing)
Shipping/Storage Pressure
50 kPa to 106 kPa
Warning:
Operate in an indoor environment only, free from moisture, flammable liquids, gases, corrosive
substances, strong electrical or magnetic fields and equipment that generates high frequency waves.
Ultrasonix cannot guarantee the proper performance of the system if used in the above-listed conditions.
Operating Temperature
32º to 104º F (0º to 40º C)
Storage Temperature
-40º to 140º F (-40º to 60º C)
Humidity
5% to 95% relative humidity, non-condensing
Light Levels
Up to 4842 Lux (foot-candles)
Shock
Designed to withstand 1.5 m (5') drops
Contaminants
Sealed to resist airborne particulate contaminants
Ventilation None
required
Operating Temperature
32º to 104º F (0º to 40º C)
Storage Temperature
-4º to 158º F (-20º to 70º C)
Humidity
80% maximum, non-condensing
Note:
For more details on the wireless adapter and other peripherals, refer to the manufacturer's User's Guides
included with the system.
Summary of Contents for SonixTouch Q+
Page 1: ...Analogic Ultrasound Service Manual SonixTouch Q Ultrasound System...
Page 2: ......
Page 4: ......
Page 14: ...Chapter 2 System Specifications 00 053 204 Revision A SonixTouch Q Service Manual 2 2...
Page 42: ...Chapter 4 Performance Testing 00 053 204 Revision A SonixTouch Q Service Manual 4 4...
Page 120: ...Chapter 8 DICOM 00 053 204 Revision A SonixTouch Q Service Manual 8 24...
Page 147: ......