SonixTouch Q+ Service Manual
00.053.204, Revision A
Chapter 8: DICOM
8-5
8.1.2.3
DICOM Storage Settings – Global Storage Settings
The
DICOM Global Storage Settings
dialog specifies global image storage parameters.
Figure 8-5: DICOM Storage Settings – Global Storage Settings
Table 8-4: DICOM Storage Settings – Global Storage Settings
St
o
ra
g
e
O
p
ti
o
n
s
Schedule
Select an auto-transfer setting:
End of Exam
,
Immediate
,
On Idle
.
Note:
This applies only if:
• at least one device is enabled
• a
Custom Key
has been configured for
DICOM Store
Max Frames per
Compress Loop
Maximum number of frames compressed at one time: 10–300. The default is 50.
Note:
Once all frames in a given file are compressed, the completed file will be transferred.
Compression type is based on the
Multi-Frame
format selected in the
Storage
Settings
dialog.
F
ailure H
and
ling
Show Error Balloons
Select to enable the display of
DICOM Storage
error messages (e.g.,
Failed to connect
to DICOM
).
Limi
t At
temp
ts
Select to configure error handling after a failed
DICOM
storage attempt.
Note:
If
Limit Attempts
is not selected, the following four options will be grayed out.
Max Storage
Attempts
Select the maximum number of times the system will retry after a failed storage attempt:
1–1000. The default is 50.
Note:
Failure may occur during
DICOM
file creation or
DICOM
file transfer. The retry
setting applies to both failure types.
Create New Error
File (Days)
Select the length of time (in days) that initiates the creation of a new error file: 1–90. The
default is 30 days.
Note:
When a new file is created, the existing file is saved, renamed and stored for 180 days.
This setting works in conjunction with
Max Error File Size (Mb)
. A new file is created
based on whichever limit is reached first.
Max Error File
Size (Mb)
Select the file size (in Mb) that initiates the creation of a new error file: 1–50. The default
is 10 Mb.
Note:
When a new file is created, the existing file is saved, renamed and stored for 180 days.
This setting works in conjunction with
Create New Error File (Days)
. A new file is
created based on whichever limit is reached first.
Error File
Retention (Days)
Select the length of time (in days) that each error file will be retained: 30–180. The default
is 180 days.
Storage Folder
Enter, or
Browse
for, the location (local or remote) where the images will be stored.
Note:
If a value is specified,
DICOM
(specifically, the
AE Configuration
and
Storage
Commitment
dialogs) will be disabled, preventing the transfer of images to an
SCP
.
Insert (Symbol)
Enables the insertion of text symbol(s) not available on the keyboard (e.g., punctuation,
symbols and letters from other languages).
Summary of Contents for SonixTouch Q+
Page 1: ...Analogic Ultrasound Service Manual SonixTouch Q Ultrasound System...
Page 2: ......
Page 4: ......
Page 14: ...Chapter 2 System Specifications 00 053 204 Revision A SonixTouch Q Service Manual 2 2...
Page 42: ...Chapter 4 Performance Testing 00 053 204 Revision A SonixTouch Q Service Manual 4 4...
Page 120: ...Chapter 8 DICOM 00 053 204 Revision A SonixTouch Q Service Manual 8 24...
Page 147: ......