Chapter 7: Field Service Components
00.053.204, Revision A
SonixTouch Q+ Service Manual
7-28
5.
From the back, gently push the modulo forward until the "Service Access" screws are revealed.
6.
Using the #2 Phillips screwdriver, unfasten the two screws securing the Front Block.
7.
Swing open the Front Block/Ultrasound Module to expose the interior.
Caution:
When pushing the modulo forward, take care to ensure that the cabling is protected from nicks and
cuts.
Note:
The screws will be labeled with a
Service Access
sticker.
Caution:
Only open the door far enough to view inside and locate the cable tie securing the PCI cables to the
inside of the door.