32
MAINTENANCE
EMPTYING AND CLEANING THE IN-LINE WATER TRAP
The water trap should be checked and emptied on a regular basis. Water vapor will condense in the probe
line, which may cause the water trap to fill beyond the maximum level mark. If this occurs water could be
drawn into the analyzer causing damage. Care should be taken at all times to avoid this risk.
Emptying of the water trap is detailed below:
Carefully remove the end cap from the in-line housing. Dispose of the condensate in a suitable drain; care
must be taken, as it could be acidic. If condensate spills onto the skin or clothing, clean off immediately
using fresh water and seek medical advice if problems occur.
CHANGING THE PARTICLE FILTER
This is a very important part of the analyzer and should be changed regularly. It prevents dust and dirty
particles entering the pump and sensors and hence causing damage. The filter MUST be changed when it
appears discolored.
Remove the end cap from the filter housing. Carefully remove the paper filter element and dispose of it.
Clean the inside of the filter housing with a suitable soft cloth. Insert a new filter element onto the spigot
on the filter end cap and carefully insert it into the filter body.