Fit a new ribbon guiding the paper roll between the exposed ribbon and cartridge body.
PROBLEM SOLVING
The following is a list of problems that may occur on the instrument through its operating life. If the
cause of the fault is not easy to identify then we advise you to contact the Kane International Service
Department or an International Distributor for expert advice.
Fault symptom
Causes
Oxygen too high
CO2 too low
Air leaking into probe, tubing, water trap,
connectors or internal to instrument.
Oxygen cell needs replacing.
Analyzer not holding charge
Analyzer not charging
Battery exhausted.
AC charger not giving correct output.
Fuse blown in charger plug.
Analyzer does not respond to flue gas
Particle filter blocked.
Probe or tubing blocked.
Pump not working or damaged with contaminants.
Flue temperature readings erratic
Temperature plug reversed in socket.
Faulty connection or break in cable or plug.
Analyzer automatically switches off in operation. Battery below alarm level.
Battery quickly discharging and is faulty.
Display is blank.
The contrast setting has been lost and requires
resetting. Disconnect handset lead and reconnect.
Set contrast as in MENU : SCREEN : CONTRAST