PRODUCT SPECIFICATION
UNIT
Temp Measurement
Resolution
Accuracy
Range
Flue Temperature
0.1° (C/F)
1.0°C ± 0.3% of reading
0 - 1100°C
32 - 2140°F
* Use high temperature probe
for gases >600°C/1112°F
Inlet Temperature
0.1° (C/F)
1.0°C ± 0.3% of reading
0 - 600°C
0 - 999°F
Gas Measurement
*1
Resolution
Accuracy
Range
Oxygen (02):
0.1%
-0.1% +0.2% 02
0- 25%
Carbon monoxide (CO):
(standard: H compensated)
1ppm
± 5ppm < 100ppm
5% of reading < 2000ppm
± 10% of reading >2000ppm
0 - 10,000ppm
Nitric oxide (NO):
(high range)
1ppm
± 5ppm < 100ppm
± 5% of reading > 100ppm
0 - 5,000ppm
Nitric oxide (NO):
(low range)
1ppm
± 2ppm < 30ppm
± 5ppm > 30ppm
0 - 100ppm
Nitrogen dioxide (NO2):
1ppm
± 5ppm < 100ppm
± 10ppm < 500ppm
± 5% of reading > 500ppm
0 - 1,000ppm
Sulphur dioxide (SO2)
(low range):
1ppm
± 2ppm < 30ppm
± 5ppm > 30ppm
0 - 100ppm
Sulphur dioxide (SO2)
(high range):
1ppm
± 5ppm < 100ppm
± 5% of reading > 100ppm
0 - 5,000ppm
Hydrogen sulphide:
1 ppm
± 5ppm < 100ppm
± 5% of reading > 100ppm
0 - 1,000ppm
Pressure:
0.01mbar
± 0.5% Full scale
0 - 150mbar
Carbon dioxide (CO2)
*2
:
0.1%
± 0.3% CO2
0 - Fuel Value
Efficiency
*2
:
0.1%
± 1%
0 - 100%
*1
using dry test gases at STP
*2
calculated
NB: all ppm reading can be displayed in mg/m3 and can be normalized
Summary of Contents for K9206