PRESSURE:
•
Allows the selection of pressure units.
•
Select from: mbar, In H2O, mm H2O, hPa, psi.
•
Use UP or DOWN keys to scroll through the selection. Press ENTER to select.
SET PERCENTAGE REFERENCE O2:
•
Can be set between 0% (equivalent to OFF) and 10%.
•
Use UP or DOWN keys to scroll through the selection. Press ENTER to select.
SET NOx CALCULATION:
• Select from: SUM, NO2 or NO
• Set REFERENCE NOx to the percentage required or defined by local regulations.
Typically 5% NO2 is added to an NO reading. The value can be user set.
• SUM adds the readings from an NO sensor and an NO2 sensor when fitted
• NO calculates an NOx reading from the NO reading where NOx = NO x 1.1
NO2 calculates an NOx reading from the NO2 reading where NOx = NO2 x 2.05
• Use UP or DOWN keys to scroll through the selection. Press ENTER to select.
CONVERSION FACTORS:
Display the Propane Equivalency Factor PEF and the Methane Equivalency Factor for the IR module
(if fitted) and the pitot factor – change from 0.10 to 1.00 as determined by the pitot tube being used.
CO ALARM: